9F3S | pdb_00009f3s

13pf mosaic 20%E254Q - 80% E254N microtubule from recombinant human tubulin decorated with EB3


ELECTRON MICROSCOPY

Starting Model(s)

Initial Refinement Model(s)
TypeSourceAccession CodeDetails
experimental modelPDB 7SJ7 

Refinement

RMS Deviations
KeyRefinement Restraint Deviation
f_dihedral_angle_d11.6482
f_angle_d0.5196
f_chiral_restr0.0413
f_plane_restr0.0041
f_bond_d0.0022
Sample
13pf E254Q microtubule from recombinant human tubulin decorated with EB3
Specimen Preparation
Sample Aggregation StateHELICAL ARRAY
Vitrification InstrumentFEI VITROBOT MARK IV
Cryogen NameETHANE
Sample Vitrification Details
3D Reconstruction
Reconstruction MethodHELICAL
Number of Particles30010
Reported Resolution (Å)4.2
Resolution MethodFSC 0.143 CUT-OFF
Other Details
Refinement Type
Symmetry TypeHELICAL
Axial SymmetryC13
Axial Rise9.58598
Angular Rotation-27.6855
Map-Model Fitting and Refinement
Id1 (7sj7)
Refinement SpaceREAL
Refinement ProtocolRIGID BODY FIT
Refinement Target
Overall B Value150
Fitting Procedure
Details
Data Acquisition
Detector TypeFEI FALCON IV (4k x 4k)
Electron Dose (electrons/Å**2)7.27
Imaging Experiment1
Date of Experiment
Temperature (Kelvin)
Microscope ModelTFS GLACIOS
Minimum Defocus (nm)1000
Maximum Defocus (nm)2500
Minimum Tilt Angle (degrees)
Maximum Tilt Angle (degrees)
Nominal CS
Imaging ModeBRIGHT FIELD
Specimen Holder Model
Nominal Magnification
Calibrated Magnification
SourceFIELD EMISSION GUN
Acceleration Voltage (kV)200
Imaging Details
EM Software
TaskSoftware PackageVersion
PARTICLE SELECTIONRELION3.1.2
CTF CORRECTIONRELION3.1.2
MODEL FITTINGCoot0.9.8.92
CLASSIFICATIONRELION3.1.2
RECONSTRUCTIONPHENIX1.21
MODEL REFINEMENTPHENIX1.21
Image Processing
CTF Correction TypeCTF Correction DetailsNumber of Particles SelectedParticle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION54200