ELECTRON MICROSCOPY
| Sample |
|---|
| Skeletal muscle actin |
| Specimen Preparation | |
|---|---|
| Sample Aggregation State | FILAMENT |
| Vitrification Instrument | FEI VITROBOT MARK IV |
| Cryogen Name | ETHANE |
| Sample Vitrification Details | Plunged into liquid ethane (FEI VITROBOT MARK IV) |
| 3D Reconstruction | |
|---|---|
| Reconstruction Method | HELICAL |
| Number of Particles | |
| Reported Resolution (Å) | 4.7 |
| Resolution Method | FSC |
| Other Details | (Helical Details: IHRSR) |
| Refinement Type | |
| Symmetry Type | HELICAL |
| Axial Symmetry | C1 |
| Axial Rise | 27.6 |
| Angular Rotation | 166.7 |
| Data Acquisition | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| Detector Type | FEI FALCON II (4k x 4k) | ||||||||
| Electron Dose (electrons/Å**2) | |||||||||
| Imaging Experiment | 1 |
|---|---|
| Date of Experiment | 2014-10-09 |
| Temperature (Kelvin) | |
| Microscope Model | FEI TITAN KRIOS |
| Minimum Defocus (nm) | 500 |
| Maximum Defocus (nm) | 3000 |
| Minimum Tilt Angle (degrees) | |
| Maximum Tilt Angle (degrees) | |
| Nominal CS | 2.7 |
| Imaging Mode | BRIGHT FIELD |
| Specimen Holder Model | FEI TITAN KRIOS AUTOGRID HOLDER |
| Nominal Magnification | |
| Calibrated Magnification | |
| Source | FIELD EMISSION GUN |
| Acceleration Voltage (kV) | 300 |
| Imaging Details |
| EM Software | ||
|---|---|---|
| Task | Software Package | Version |
| RECONSTRUCTION | EMAN | 2 |
| RECONSTRUCTION | SPIDER | |